Power Calibration    Related Topics

The Power Calibration submenu configures and starts a source or receiver power calibration.

In true differential mode only the source power calibration results are used, no receiver power calibration is possible.

Power correction data can be stored to a cal group and applied to several channels; see description of the Calibration Manager.
In remote control it is possible to read, modify and reload power calibration data. Refer to the description of the commands SOURce<Ch>:POWer<Pt>:CORRection:DATA and [SENSe<Ch>:]CORRection:POWer<port_no>:DATA.

Remote control:

SOURce<Ch>:POWer:CORRection...
[SENSe<Ch>:]CORRection:POWer...


Source Power Cal...

Opens the configuration dialog for a source power calibration.

Purpose of a source power calibration

A source power calibration ensures accurate power levels of the generated wave at an arbitrary reference plane in the measurement path. The calibrated reference plane is typically the input of the DUT.

In a frequency sweep, the power at the reference plane is maintained at a constant Cal Power value. The source power calibration eliminates frequency response errors in the signal path between the source and the reference plane. It is possible to introduce an arbitrary attenuation or gain into the signal path so that the Cal Power is not restricted to the power range of the source. A typical application for a power calibration in a frequency sweep is the measurement of the gain of an amplifier across a frequency range but at a fixed input power.

In a power sweep, the power calibration ensures that the power at the reference plane is either constant or a linear function of the stimulus power. A typical application for a power calibration in a power sweep is the measurement of the gain of an amplifier across a power range but at a fixed frequency.   

The correction data acquired in a frequency or power sweep is re-used if a Time or CW Mode sweep is activated.

Calibration procedure

The source power calibration requires an external power meter, to be connected via GPIB bus, USB or LAN interface. Use the USB-to-IEC/IEEE Adapter (option R&S ZVAB-B44) to control devices equipped with a GPIB interface. The power sensor can be connected directly at the reference plane or to any other point in the test setup where the signal power is known to be proportional to the power at the reference plane.  

The source power calibration involves several steps:

  1. The analyzer performs a first calibration sweep at the source power that is likely to produce the target power (Cal Power) at the reference plane. A known attenuation or gain at the source port and in the signal path between the source port and the reference plane is taken into account.

  1. The power measured at the reference plane is displayed in the calibration sweep diagram and compared to the Cal Power. The comparison serves as an input for the correction of the source power.

  2. Steps 1 and 2 are repeated for a specified Number of Readings or after the deviation between the measured power at the reference plane and the Cal Power is below a specified Tolerance.

  3. After the last calibration sweep, it is possible to perform an additional Verification Sweep in order to check the accuracy.

After a power calibration one can expect the power at the reference plane to be within the range of uncertainty of the power meter. After a change of the sweep points or sweep range, the analyzer interpolates or extrapolates the calibration data.

The Source Power Cal dialog provides the following control elements:

Remote control:

SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire]
SOURce<Ch>:POWer<Pt>:CORRection:COLLect[:ACQuire]

SOURce<Ch>:POWer<Pt>:CORRection[:STATe]

SOURce<Ch>:POWer<Pt>:CORRection:GENerator[:STATe]

SOURce<Ch>:POWer<Pt>:CORRection:PMETer:ID

SOURce<Ch>:POWer<Pt>:CORRection:COLLect[:ACQuire]:VERification[:STATe]

SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire]:VERification:RESult?


Calibration Sweep Diagram

The calibration sweep diagram displays the power measured by the external power meter while a source power calibration is performed.

The diagram is automatically scaled so that the center corresponds to the selected Cal Power. For frequency sweeps, a horizontal trace close to the Cal Power indicates an accurate power calibration. For power sweeps, the ideal trace is a power ramp.

Below the sweep diagram, the status messages Calibration in Progress/Finished and Verification in Progress/Finished show the progress of the calibration procedure.


Modify Source Power Cal Settings

The Modify Source Power Cal Settings dialog displays the target power for the source power calibration and defines stop criteria.

The dialog provides the following control elements:

The analyzer performs either a Flatness Cal or a Reference Receiver Cal, or both. The two options must not be disabled simultaneously.

Conditions for fast power calibration

To use fast power calibration, ensure that the analyzer is configured as follows:

  1. An internal source is used.

  2. The Reference Receiver Cal is enabled.

  3. On an R&S ZVA network analyzer that is equipped with option R&S ZVA<n>-B16, Direct Generator/Receiver Access, the jumper for the REF OUT / REF IN connectors is connected so that the reference receiver loop is closed.

Remote control:

SOURce<Ch>:POWer<Pt>:CORRection:NREadings
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:AVERage[:COUNt]

SOURce<Ch>:POWer<Pt>:CORRection:COLLect:AVERage:NTOLerance

SOURce<Ch>:POWer<Pt>:CORRection:COLLect:CFACtor

SOURce<Ch>:POWer<Pt>:CORRection:COLLect:FLATness

SOURce<Ch>:POWer<Pt>:CORRection:COLLect:RRECeiver

SOURCe<Ch>:POWer<Pt>:CORRection:FAST

SOURce<Ch>:POWer<Pt>:CORRection:COLLect:METHod

SOURce<Ch>:POWer<Pt>:CORRection:COLLect:PMReadings

SOURce<Ch>:POWer<Pt>:CORRection:OSOurces[:STATe]


Power Meter Settings

The Power Meter settings in the Source Power Cal dialog (and similar dialogs) are used to select the external power meter that is used to measure the exact source power at the calibration point and to ensure maximum accuracy of the power measurement. Power meters (Pmtr1, Pmtr2, ...) must be configured in the System Configuration External Power Meters dialog before they appear in the list.

The Power Meter panel provides the following buttons.

Zeroing

Zeroing calibrates the external power meter by adjusting its reading at zero signal power. For this purpose, the RF cable between the analyzer and the power sensor must be disconnected (see tips below!). R&S power sensors and power meters automatically detect the presence of any significant input power. This aborts zeroing and generates an error message. Zeroing can take a few seconds, depending on the power meter model; refer to the documentation of your external power meter for more information.

Repeat zeroing:

A reset of the network analyzer does not affect the last zeroing result.


Power Meter Correction

Defines the frequency range for the sensitivity correction of the power meter and an additional transmission coefficient in front of the power meter or the DUT.  

The settings appear in two panels:

Test setups with "additional two-port" devices

The two test setups for additional two-ports can be depicted as follows.

Case A: 2-port in front of power sensor (during calibration)

Test and measurement procedure:

1. Perform the calibration with the additional two-port between the analyzer port and the power sensor. During the calibration the analyzer increases the power sensor values by the 2-port transmission coefficients. The calibrated reference plane corresponds to the input of the additional 2-port.

2. Perform the measurement without the additional two-port.   

Practical example:

Case B: 2-port in front of DUT (during measurement)

Test and measurement procedure:

1. Perform the calibration without the additional two-port. During the calibration the analyzer decreases the power sensor values by the 2-port transmission coefficients. The calibrated reference plane corresponds to the output of an additional 2-port which is placed in-between the NWA port and the DUT.

2. Perform the measurement with the additional two-port.   

Practical example:

Remote control:

[SENSe<Ch>:]PMMO ON | OFF
SOURce<Ch>:POWer:CORRection:TCOefficient[:STATe]

SOURce<Ch>:POWer:CORRection:TCOefficient:CALibration

SOURce<Ch>:POWer:CORRection:TCOefficient:INSert

SOURce<Ch>:POWer:CORRection:TCOefficient:DEFine

SOURce<Ch>:POWer:CORRection:TCOefficient:COUNt?

SOURce<Ch>:POWer:CORRection:TCOefficient:DELete[:DUMMy]

SOURce<Ch>:POWer:CORRection:TCOefficient:DELete:ALL

SOURce<Ch>:POWer:CORRection:TCOefficient:FEED

MMEMory:LOAD:CORRection:TCOefficient "<file_name>"[,"parameter"]

MMEMory:SAVE:CORRection:TCOefficient "<file_name>"


Receiver Power Cal...

Opens the configuration dialog for a receiver power calibration. The receiver power calibration data is acquired in a calibration sweep after appropriate selection of the calibrated wave quantity and the source.

Purpose of a receiver power calibration

A receiver power calibration ensures that the power readings at a specified input port of the analyzer agree with the source power level calibrated at an arbitrary reference plane. Typically, the reference plane is at the input of the receiver so that the calibration eliminates frequency response errors in the calibrated receiver.

A receiver power calibration generally improves the accuracy of power (wave quantity) measurements.  

The correction data acquired in a frequency or power sweep is re-used if a Time or CW Mode sweep is activated.

Calibration procedure and calibration sweep

The receiver power calibration is based on a wave with known power.

The receiver power calibration involves a single calibration sweep. The calibration sweep is performed with current channel settings but with a maximum IF bandwidth of 1 kHz. Smaller IF bandwidths are maintained during the calibration sweep; larger bandwidths are restored after the sweep.

The analyzer measures the power at each sweep point, compares the result with the nominal power of the source, and compiles a correction table.

An acoustic signal indicates the end of the calibration sweep. At the same time, a message next to the calibrated source indicates the status of the receiver power calibration. After a change of the sweep points or sweep range, the analyzer interpolates or extrapolates the calibration data.  

The Receiver Power Cal dialog provides the following control elements:

The accuracy of the receiver power calibration depends on the accuracy of the source. Depending on the test setup and reference plane, it is recommendable to perform a source power calibration before the receiver power calibration.  

Remote control:

[SENSe<Ch>:]CORRection:POWer:ACQuire
[SENSe<Ch>:]CORRection:POWer:AWAVe[:STATe]

[SENSe<Ch>:]CORRection:POWer:DATA

[SENSe<Ch>:]CORRection:POWer[:STATe]


Power Calibration Labels

The status of a power calibration is indicated in the trace list and in the Used Source panel of the Receiver Power Cal dialog. The following messages can be displayed:

A lower label in the list has priority over the higher labels (e.g. if the power calibration is interpolated and the source power is changed, then the label PCao is displayed).

The labels appear in the trace list, provided that the measurement result is an S-parameter, wave quantity, ratio, or a derived quantity (stability factor, PAE), and that the effects of an inaccurate source or receiver power have been consistently removed. The conditions for consistency depend on the measured quantity (see also Data Flow):

Remote control:

[SENSe<Ch>:]CORRection:PSTATe?


Interpolation and Extrapolation

The analyzer can interpolate and extrapolate power correction data so that a source or receiver power calibration can be reused after a change of the frequency sweep range:  


Power Correction On

Applies all available source and/or receiver power calibrations to the active channel. This function is disabled if no power correction data has been acquired for the active channel.  

Uncheck the ...Power Correction Off boxes in the Source Power Cal and Receiver Power Cal dialogs to enable an individual source or receiver calibration. Press Power Correction Off to disable the power calibration of all sources and receivers.

Remote control:

SOURce<Ch>:POWer:CORRection:STATe ON
[SENSe<Ch>:]CORRection:POWer:STATe ON


Power Correction Off

Disables all source and receiver power calibrations of the active channel so that all power correction data is ignored. This function is disabled if no power correction data has been acquired for the active channel.

Check the ...Power Correction Off boxes in the Source Power Cal and Receiver Power Cal dialogs to disable an individual source or receiver calibration. Press Power Correction On to re-enable the power calibration of all sources and receivers.

Remote control:

SOURce<Ch>:POWer:CORRection:STATe OFF
[SENSe<Ch>:]CORRection:POWer:STATe OFF