SOURce<Ch>:POWer<Pt:CORRection:TCOefficient...    Related Topics  

This subsystem defines the transmission coefficients for an additional two-port in the test setup (power loss list)).

 

Command tree: SOURce<Ch>POWer<Pt>:CORRection:TCOefficient

:CALibration

:COUNt?

:DEFine

:DELete

:ALL

[:DUMMy]

:FEED

:INSert

[:STATe]


SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:CALibration <Boolean>

Selects the position of the additional two-port in the test setup.

<Ch>

Calibrated channel number. This parameter is ignored; the transmission coefficient settings are valid for all channels.  

 

<Pt>

Calibrated port number. This parameter is ignored; the transmission coefficient settings are valid for all sources.

 

<Boolean>

ON – Power sensor (during calibration)
OFF – Device under test (during measurement)

*RST value

ON

 

SCPI, Command Types

Device-specific, command or query

 

Example:

*RST; :SOUR:POW:CORR:TCO:CAL OFF

Select the test setup with the additional two-port in front of the DUT.

SOUR:POW:CORR:TCO:DEF 1GHz, -5; DEF 2GHz, -10; DEF2?  

Define two points in the power loss list; query the second point.  

SOUR:POW:CORR:TCO:INS2 1.5 GHz, -7.5

Insert a third point as point no. 2 in the list.  

SOUR:POW:CORR:TCO:COUN?

Query the number of points. The response is 3.  

SOUR:POW:CORR:TCO ON

Enable the use of two-port  transmission coefficients.  

SOUR:POW:CORR:TCO:FEED 'Trc1'

Replace the previous 3 points by the trace points of the default trace 'Trc1'.   

MMEM:STOR:CORR:TCO 'C:\Rohde&Schwarz\NWA\PowermeterCorrLists\Test.pmcl'  

Store the power loss list to the power meter correction list file Test.pmcl in the directory C:\Rohde&Schwarz\NWA\PowermeterCorrLists.  

MMEM:LOAD:CORR:TCO 'C:\Rohde&Schwarz\NWA\PowermeterCorrLists\Test.pmcl'  

Re-load the stored power meter correction list file.  

SOUR:POW:CORR:TCO:DEL1  

Delete the first point in the list.  

SOUR:POW:CORR:TCO:DEL:ALL; :SOUR:POW:CORR:TCO?

Clear the entire list. Query whether the transmission coefficients are still taken into account. The response is 0.  

MMEM:LOAD:CORR:TCO 'C:\Rohde&Schwarz\NWA\Traces\Test.s2p', 'S21'  

Load a power loss list from a previously created 2-port Touchstone file. .  

 

 


SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:COUNt?

Queries the number of frequency values and transmission coefficients in the power loss list.

<Ch>

Calibrated channel number. This parameter is ignored; the transmission coefficient settings are valid for all channels.  

 

<Pt>

Calibrated port number. This parameter is ignored; the transmission coefficient settings are valid for all sources.

 

SCPI, Command Types

Device-specific, query only

 

Example:

See SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:CALibration

 

 


SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:DEFine<List_No> <frequency>, <transmission>

Adds a new frequency and transmission coefficient to the end of the power loss list.

<Ch>

Calibrated channel number. This parameter is ignored; the transmission coefficient settings are valid for all channels.  

 

<Pt>

Calibrated port number. This parameter is ignored; the transmission coefficient settings are valid for all sources.

 

<List_no>

Number of point in the list. This suffix is only used for queries.  

<frequency>

Stimulus frequency value

Range [def. unit]

According to frequency range of the analyzer; see list of analyzer models [Hz]. If several points with identical frequencies are added, the analyzer automatically ensures a frequency spacing of 1 Hz.  

<transmission>

Transmission coefficient

Range [def. unit]

–300 dB to +200 dB [dB]

*RST values

n/a

 

SCPI, Command Types

Device-specific, command or query. The query returns the frequency and transmission coefficient no. <List_No>.

 

Example:

See SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:CALibration

 

 


SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:DELete<List_No>:ALL

Clears the power loss list.  

<Ch>

Calibrated channel number. This parameter is ignored; the transmission coefficient settings are valid for all channels.  

 

<Pt>

Calibrated port number. This parameter is ignored; the transmission coefficient settings are valid for all sources.

 

<List_no>

Number of point in the list. This suffix is ignored; the command deletes all points.  

*RST values

n/a

 

SCPI, Command Types

Device-specific, no query.  

 

Example:

See SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:CALibration

 

 


SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:DELete<List_No>[:DUMMy]

Deletes a single point no. <List_No> in the power loss list.  

<Ch>

Calibrated channel number. This parameter is ignored; the transmission coefficient settings are valid for all channels.  

 

<Pt>

Calibrated port number. This parameter is ignored; the transmission coefficient settings are valid for all sources.

 

<List_no>

Number of point in the list.  

*RST values

n/a

 

SCPI, Command Types

Device-specific, no query.  

 

Example:

See SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:CALibration

 

 


SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:FEED '<trace_name>'

Selects a trace which provides the points for the power loss list..

<Ch>

Calibrated channel number. This parameter is ignored; the transmission coefficient settings are valid for all channels.  

 

<Pt>

Calibrated port number. This parameter is ignored; the transmission coefficient settings are valid for all sources.

 

'<trace_name>'

String parameter for the trace name, e.g. 'Trc1'. The trace must exist in the active setup.

*RST value

n/a

 

SCPI, Command Types

Device-specific, no query

 

Example:

See SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:CALibration

 

 


SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:INSert<List_No> <frequency>, <transmission>

Inserts a new frequency and transmission coefficient at position no. <List_no> in the power loss list. The following points are shifted down in the list.  

<Ch>

Calibrated channel number. This parameter is ignored; the transmission coefficient settings are valid for all channels.  

 

<Pt>

Calibrated port number. This parameter is ignored; the transmission coefficient settings are valid for all sources.

 

<List_no>

Number of inserted point in the list.  

<frequency>

Stimulus frequency value

Range [def. unit]

According to frequency range of the analyzer; see list of analyzer models [Hz]. If several points with identical frequencies are added, the analyzer automatically ensures a frequency spacing of 1 Hz.  

<transmission>

Transmission coefficient

Range [def. unit]

–300 dB to +200 dB [dB]

*RST values

n/a

 

SCPI, Command Types

Device-specific, command or query.  

 

Example:

See SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:CALibration

 

 


SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient[:STATe] <Boolean>

Enables or disables the use of two-port transmission coefficients.

The power loss list must contain at least one point before the transmission coefficients can be enabled.

<Ch>

Calibrated channel number. This parameter is ignored; the transmission coefficient settings are valid for all channels.

 

<Pt>

Calibrated port number. This parameter is ignored; the transmission coefficient settings are valid for all sources.

 

<Boolean>

Enables (ON) or disables (OFF) the two-port transmission factors.

*RST value

OFF [–]

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See SOURce<Ch>:POWer<Pt>:CORRection:TCOefficient:CALibration