Converter Control (Option R&S ZVA-K8)   Related Topics

Measurements at frequencies beyond the analyzer's operating range (mm-wave measurements) are achieved by combining a frequency-converting measurement with an external test set (frequency converter). Rohde & Schwarz offers frequency converter types for various frequency ranges. E.g. the frequency converters R&S ZVA-Z110 and R&S ZV-Z110E extend the frequency range of R&S ZVA or R&S ZVT network analyzers to a range between 75 GHz and 110 GHz.

The test ports of the frequency converters and the connecting elements are rectangular waveguides. For the R&S ZVA-Z110(E) converter, the dimensions of the waveguides are according to EIA WR-10, equivalent to RCSR WG-27. Refer to the Quick Start Guide and the "Specifications" supplied with your converter for detailed information about the frequency range and the waveguide type of your converter model.  

Two different converter types are available:

Required equipment and options

Frequency extension is available on ZVA and ZVT analyzers which are specified for frequencies up to 20 GHz or above (e.g. ZVT 20, ZVA24, ZVA 40). The required firmware version depends on the converter mode. Refer to the release notes (What's new) or to the Quick Start Guide of your converter.  

The following additional options and accessories are needed:


Test Setup and Principle of the Measurement

The frequency converters use frequency multipliers to transform the RF source signal from one of the network analyzer ports into a high-frequency stimulus signal. A dual directional coupler separates the reference and measurement channels from the waveguide test port. A second signal (Local Oscillator, LO) is used for down-conversion of the reference and measurement channels. The LO signal can be provided either by a second analyzer port or by an external generator. The down-converted signals are fed to the REF IN and MEAS IN input connectors of the analyzer port providing the RF source signal.

The schematic test setup for a two-port transmission measurement using four analyzer ports (no external generator) is shown below.

A measurement using converters without electronic attenuator involves the following stages:

  1. Selection of the converter type and of the test setup in the Frequency Converter dialog

  2. Connection of the frequency converters

  3. System Error Correction using a suitable waveguide calibration kit

  4. Connection of the DUT and measurement

For converters with electronic attenuators, the following additional stages are required:


Frequency Converter

The Frequency Converter tab belongs to the System Configuration menu (System System Config). It selects the frequency converter type and the external test setup as well as the power transfer model (especially for converters with electronic attenuators).  

The dialog provides the following control elements:  

The frequency-converter settings belong to the global resources; they are not changed upon a Preset or *RST.

Analyzer settings with active frequency converter

After a particular converter type and test setup is applied, the analyzer is automatically configured as follows:

All signal settings appear in the Port Configuration dialog as soon as the frequency converter mode is active. In the example below, a sweep range between 75 GHz and 110 GHz is selected. The frequency multipliers for the RF signals have a multiplication factor of 6, so the actual output frequency at the RF ports 1 and 2 is between 12.5 GHz and 18.3333333 GHz. The frequency of the LO signal is such that the reference and measurement channels are received at the fixed IF frequency of 279 MHz (shown in the Receiver section of the Port Configuration dialog). The LO frequency is always below the RF frequency (upper sideband).

It is also possible to define port-specific frequencies and powers at the converter ports; see Power and Frequency Settings.  

Remote control:

[SENSe<Ch>:]FREQuency:CONVersion:DEVice:MODE
[SENSe<Ch>:]FREQuency:CONVersion:DEVice:NAME


Power Transfer Models

The Power Transfer (All Converter Ports) settings in the Frequency Converter tab define how the analyzer controls the output power at all converter ports. The general settings can be changed for individual converter ports (Channel – Mode – Port Configuration – Source Power – Converter Port <n> Power – Power Transfer).

Changing the power transfer settings in the Frequency Converter tab will overwrite the individual port settings.

The basic power transfer model (Use Advanced Model off) is simple to configure and suitable for most standard applications.

The advanced model may be based on a Data Set or on a Linear Approximation of the relation between the converter input power and the converter output power.

The following table shows an overview of the different modes.

Power Transfer Model

Converters with electronic attenuators
(R&S ZVA-ZxxxE)
Converter output power controlled by...

Converters without electronic attenuators
(R&S ZVA-Zxxx)
Converter output power controlled by...

Basic

Electronic waveguide attenuators in the converters

(Power adjustment screw)

Advanced, data set

Measured set of attenuation values (power calibration data)

Not supported

Advanced, linear approximation

Combination of RF input power (network analyzer) and electronic waveguide attenuators (converter)

Mechanical attenuator setting (Power Transfer dialog) and power adjustment screw

Remote control:

[SENSe<Ch>:]CONVerter:AMODel
[SENSe<Ch>:]CONVerter:DESCription

[SENSe<Ch>:]CONVerter:PATH


Power Coefficients (for Converters with Electronic Attenuators)

The Power Coefficients dialog is used to enter the power coefficients of converters with external attenuators. It can be opened from the System Configuration – Frequency Converter tab after the converter is activated.

For accurate control of the converter output power, the R&S ZVA analyzer must know the (non-linear) current-power characteristic of the converters. The characteristic is sufficiently described by a third-order polynomial. A label with the four polynomial coefficients c0, c1, c2, and c3 is affixed to each converter.

Remote control:

[SENSe<Ch>:]FREQuency:CONVersion:DEVice:PCOefficient<Port>
[SENSe<Ch>:]FREQuency:CONVersion:DEVice:PCOefficient<Port>:DEFault


Connecting the Frequency Converters

Each frequency converter must be connected to the DUT, the network analyzer, and the power supply. The DUT is screwed on the waveguide flange at the front of the converter. The remaining connectors are located on the rear panel of the converter:

 

Attention!

Input powers RF IN and LO IN

The RF input power at the connectors RF IN and LO IN must not exceed the maximum values quoted in the data sheet. The maximum values are below the maximum RF source power of the network analyzer. The frequency converter mode ensures compatible source powers.

Before you connect your converter to the network analyzer, always activate the frequency converter mode using the Frequency Converter dialog and select the proper converter type and connecting diagram.

 

Connection to the network analyzer comprises the following RF input and output signals:

  1. RF source signal (input): Connect port 1 or port 2 of the analyzer to RF IN of the converters.

  2. LO signal (input): Connect port 3 or port 4 of the analyzer to LO IN of the converters. Alternative: Connect an external generator signal to LO IN; see Frequency Converter dialog.

  3. Reference signal (output): Connect REF OUT of the converters to the REF IN connector of the analyzer port providing the RF source signal. The reference signal corresponds to the incident wave (a-wave) applied to the input of the DUT.

  4. Measurement signal (output): Connect MEAS OUT of the converters to the MEAS IN connector of the analyzer port providing the RF source signal. The measured signal corresponds to the b-wave caused due to reflection or transmission at the DUT.  

For converters with electronic attenuators, an additional USB control connection between the output connector of option R& ZVA-B8 (EXTATT CTRL) on the front panel of the analyzer and the 3-pin control connector at the rear panel of the converter is required. Observe the port assignment of the NWA output connector.

The input signal at the frequency converters must be relatively stable over the entire sweep range. Use high-quality cables for the connection of the RF and LO input signals. Appropriate cables for the REF OUT and MEAS OUT signals are supplied with the frequency converters.

In view of the port arrangement at the front panel of the analyzer, it is advantageous to use test ports 1 and 2 as RF sources, ports 3 and 4 as LO sources. The frequencies of the RF and LO signal must be independent, therefore do not combine a pair of ports as RF and LO sources that is supplied by a common generator (ports 1 and 2, 3 and 4; see Coupled test ports).  

To supply the frequency converter, connect the external DC power supply provided with the converter to the 9 V / 0.5 A DC input. The power supply supports input AC voltages between 100 V and 240 V and frequencies between 47 Hz and 63 Hz. A complete test setup for a 2-port transmission measurement using converters without electronic attenuators is shown below.

Frequency converters and true differential mode
To generate a true differential signal with two frequency converters, a modified test setup with three independent sources is required. See background information inBalanced Ports and Port Groups – True Diff Mode.

 

Attention!

Supply voltage and power

The input voltage and current must not exceed the maximum values according to the rear panel labeling or the data sheet.

Always use the DC power supply included in delivery to power your frequency converter.

 


Power and Frequency Settings

While the frequency converter mode is active, the Channel – Stimulus settings of the network analyzer control the frequency and power range of the converters. In addition the Channel – Mode – Port Configuration dialog shows an additional row for each converter. The Frequency and Power settings in the Source section of the dialog serve different purposes:

Example: In the example below, the frequency at the NWA Port 1 has been increased by a 50 MHz offset. The converter source frequency six times the NWA source frequency, therefore a 300 MHz offset has been entered for Converter Port 1. This ensures a correct frequency axis during the power calibration.

Remote control:

SOURce<Ch>:FREQuency<Pt>:CONVersion:ARBitrary:CFRequency
SOURce<Ch>:POWer<Pt>:CONVerter:OFFSet
 


Calibration (for Converters without Electronic Attenuators)

The output power of the frequency converter can be set manually (at the converter) only, therefore the standard source power calibration eliminating frequency response errors in the signal path between the source and the reference plane (external power meter) is not possible. A power calibration of the reference receiver (a-wave) using an external power meter, however, is possible and recommended for measurements concerning the wave quantities a and b. Proceed as follows:

  1. Ensure that the output power of the frequency converter is not attenuated (adjust the knurled knob at the top of the converter to 2 mm).

  2. Connect an appropriate external waveguide power meter to the waveguide flange and open the Channel Calibration Start Power CalSource Power Cal dialog (see Source Power Cal).  

  3. Click Modify Settings and disable Flatness Cal, leaving Reference Receiver Cal checked.

  1. Start the calibration sweep.

This power calibration procedure ensures a reasonable accuracy of the reference power readings over a wide range of converter output powers (i.e. even if the adjusting knob is used to reduce the powers).

A Receiver Power Calibration of the b-waves (without external power meter) is possible after completed power calibration of the a-wave. Moreover, you can use the advanced power transfer model with linear approximation to adjust the screw to a desired attenuation value and use this value as a correction factor for the port power.

After the power calibration procedure a system error correction is recommended. Due to the physical properties of the mm-waves and the waveguides, measurements with frequency converters require a special calibration kit for system error correction. Rohde & Schwarz offers special waveguide calibration kits for this purpose, e.g. the calibration kits R&S ZV-WR03, R&S ZV-WR10, and R&S WR-15. The standards in the calibration kits allow all one-port and two-port calibration types supported by the network analyzer except TNA. Refer to the documentation of the calibration kit for details.  

Remote control:

SOURce<Ch>:POWer<Pt>:CORRection:DATA  


Calibration (for Converters with Electronic Attenuators)

A source power calibration for a frequency converter requires an appropriate external power meter, to be connected to the converter's waveguide flange. Waveguide power meters are configured in the ordinary way using the System Configuration – External Power Meters tab.  

To perform the source power calibration, proceed as follows:

  1. Connect the waveguide power meter and open the Channel – Calibration – Start Power Cal – Source Power Cal dialog.   

  2. Select your converter and source port from the Source pull-down list (e.g. Conv 1 for a frequency converter connected to NWA port 1).

  3. Click Modify Settings and ensure that both Flatness Cal and Reference Receiver Cal are checked.

  4. If your test setup causes strong nonlinear effects, you can choose a Convergence Factor different from one.

  5. Start the calibration sweep.

To ensure an accurate source power calibration and quick convergence, use the correct power coefficients. A receiver power calibration of the b-waves (without external power meter, using the “Receiver Power Calibration” dialog) is possible after completed source power calibration.

After the power calibration procedure a system error correction is recommended. Due to the physical properties of the mm-waves and the waveguides, measurements with frequency converters require a special calibration kit for system error correction. Rohde & Schwarz offers special waveguide calibration kits for this purpose, e.g. the calibration kits R&S ZV-WR03, R&S ZV-WR10, R&S ZV-WR12, and R&S WR15. The standards in the calibration kits allow all one-port and two-port calibration types supported by the network analyzer except TNA. Refer to the documentation of the calibration kit for details.

Remote control:

SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire]
SOURce<Ch>:POWer<Pt>:CORRection:CONVerter<Con>[:STATe]

SOURce<Ch>:POWer<Pt>:CORRection:DATA
 
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:CFACtor


Measurement

After power calibration and system error correction, the mm-wave measurement can be performed like any other network analyzer measurement. The Port Configuration settings (together with the Stimulus settings) determine the sweep range of the converted signals (i.e the input and output frequencies at the DUT ports). All measured quantities (S-parameters, wave quantities, ratios etc.) and other trace settings are available.

The following restrictions hold for measurements with external frequency converters without external attenuators:

The following example shows the transmission and reflection coefficients of a bandpass filter in the frequency range between 75 GHz and 110 GHz, i.e. the measurement range of the converter R&S ZVA-Z110.