Wave Quantities

The Wave Quantities submenu selects a wave quantity to be measured and displayed.

Wave quantities

In contrast to S-parameters, wave quantities are not system error corrected. A power calibration can be applied to wave quantities; see Data Flow.  


a1 Src Port 1, b1 Src Port 1, b2 Src Port 1, a2 Src Port 2, b1 Src Port 2, b2 Src Port 2

Select the standard 2-port wave quantities a1, a2, b1, and b2 for different source ports:

The predefined wave quantities are obtained with different source ports. a1 Src Port 1, b1 Src Port 1 and b1 Src Port 2 are measured at Port 1 of the analyzer. a2 Src Port 2, b2 Src Port 1 and b2 Src Port 2 are measured at Port 2 of the analyzer.  

The analyzer can also measure wave quantities for other source ports; see More Wave Quantities.

Remote control:

CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "A1" | ...  
[SENSe<Chn>:]FUNCtion[:ON] "...
POWer:A1" etc.

Create new trace and select name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "A1" | ...
 


More Wave Quantities

Opens a dialog to define and select arbitrary wave quantities for different detectors and source ports or higher port numbers. In true differential mode, the dialog also offers balanced wave quantities.  

The notation for wave quantities and the functionality of the More Wave Quantities dialog is analogous to the definition of S-parameters.

The measurement process for external generators Gen 1, Gen2... in the Source Port list differs from the measurement process for internal source ports:

Single-ended and balanced wave quantities

If a balanced port configuration is selected and the True Differential Mode is active, the left Wave Quantity pull-down list contains the following wave types:

Primed wave quantities

In addition to the wave quantities a and b, the analyzer provides "primed" wave quantities a', b'; see Measurements at Different Frequencies.

If a mixer delay measurement is active, the unprimed wave quantities correspond to the lower tone, the primed wave quantities correspond to the upper tone.   

Remote control:

CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "A1" | "AP1"...  
[SENSe<Chn>:]FUNCtion[:ON] "...
POWer:A1" etc.

Create new trace and select name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "A1" |
"AP1"... 

CALCulate<Chn>:FORMat:WQUType POWer | VOLTage